This application note describes a fully automated analysis method for selected equine doping compounds in equine urine. A GERSTEL MultiPurpose Sampler (MPS) with Disposable Pipette Extraction (DPX) option is employed for extraction and cleanup. After gas chromatographic separation the analytes are detected by a triple-quadrupole mass spectrometer (QqQ-MS).

The method is rugged, provides an adequate cleanup of the complex sample matrix and shows good limits of detection, from below 0.1 to just under 10 ng/mL for the various analytes that are determined.

Disposable Pipette Extraction (DPX)

Disposable Pipette Extraction (DPX)

Extraction in Seconds

DPX is a fast and efficient SPE technique used for a wide rnge of applications such as drugs of abuse, therapeutic drug monitoring, comprehensive screening, pharmacology studies (NNK), as well as pesticides in fruit and vegetables. DPX is based on unique and patented SPE devices: Pipette tips that incorporate loosely contained sorbent material, which is mixed with the sample solution. Turbulent air bubble mixing creates a suspension of sorbent in the sample ensuring optimal contact and highly efficient extraction. The extraction is performed much faster than with traditional SPE techniques.